Specifications
Current Draw at 100 VAC 50 Hz
Current Draw at 120 VAC 60 Hz
Current Draw at 220 VAC 50 Hz
Current Draw at 240 VAC 50 Hz
Gas Supply Requirements
Gas Composition
Supply Pressure, Nominal
Supply Pressure, Maximum
Carrier Gas
Supply Pressure, Minimum
Flow Rate, Nominal, Test Active, Maximum
Flow Rate, Nominal, Test Active, Minimum
Gas Composition
Supply Pressure, Nominal
Test Gas
Supply Pressure, Maximum
Supply Pressure, Minimum
Flow Rate, Nominal, Test Active
Operational Capabilities
Temperature Control Range
Time to Reach a Heated Temp (from ambient)
Time to Reach a Cooled Temp (from ambient)
Humidity Control, Film Cell, Carrier Gas
Humidity Control, Film Cell, Test Gas
RH Film Cell-to-Cell Uniformity
Time to Reach Humidity Setting
10-2
Table 10.3: Electrical Specifications
Table 10.4: Gas Supply Specifications
10 °C to 40 °C ± 0.2 °C
Approximately 45 minutes to reach 99% of a 40 °C setting
Approximately 2 hours to reach 99% of a 10 °C setting
0% to 90% RH ± 3 RH
0% to 90% RH ± 3 RH
± 3 RH
Approximately 3 hours to reach 99% of RH setting
Revision L
OX-TRAN Model 2/12 R Operator's Manual
1.80 A nominal
1.50 A nominal
0.82 A nominal
0.75 A nominal
100% Nitrogen
29 PSI, (2.0 Bar), (200 kPa)
32 PSI, (2.2 Bar), (221 kPa)
26 PSI, (1.5 Bar), (1792 kPa)
600 cc/minute
60 cc/minute
100% Oxygen (purity 99.9%)
29 PSI, (2.0 Bar), (200 kPa)
32 PSI, (2.2 Bar), (221 kPa)
21 PSI, (1.38 Bar), (145 kPa)
40 cc/minute
(< 10 ppm Oxygen)
MOCON, Inc.