Permeation Test Overview; Individual Zero Phase; Conditioning The Sample; Rezero State - Ametek mocon OX-TRAN 2/12 R Operator's Manual

Oxygen transmission rate system
Table of Contents

Advertisement

OX-TRAN Model 2/12 R Operator's Manual
Introduction

Permeation Test Overview

A "Permeation Test" consists of up to two phases that are performed using a Test Cell. The "Cell" may
contain a flat film sample or be connected to a package fixture. The Test Phase (always present) is used to
measure the transmission rate of the test sample. The Individual Zero Phase is optional and may occur at
the beginning or end of a Permeation Test. Each phase in the test consists of a series of discrete states. A
Permeation Test may utilize all or only a subset of the possible test states. During a test the steps (or
states) within each phase are executed in a fixed order. This order is called the "Test Sequence". A brief
description of the phases and states within a permeation test are given below:

Individual Zero Phase

An Individual Zero Phase is used to determine the amount of oxygen that is present in the carrier gas from
factors other than actual transmission through the sample. This amount is subtracted from the transmission
rate obtained for each sample during testing. For more information on when and how to use an Individual
Zero Phase see "Individual Zero Processing" in Chapter 3.

Conditioning the Sample

Almost every material requires a period of time to acclimate to the environment and reach equilibrium. A
time period can be specified to allow the barrier to "Condition" to the test environment. The Conditioning
Period when used will occur at the beginning of each test phase.

ReZero State

Periodically the instrument baseline must be measured so that any changes in the baseline will not affect
the accuracy of your transmission rate data. This is done during a "ReZero State" by measuring the
apparent oxygen transmission rate of the ReZero Cell.

Cell Examination State

The Cell Examination State is used to measure the transmission rate of the barrier sample. The carrier gas
leaving the Test Cell is sent to the oxygen sensor where the amount of oxygen contained in the gas is
measured to determine a transmission rate.

Bypass State

When no gas is being routed to the oxygen sensor it is said to be in the Bypass State. The Bypass State is
used to maximize the useful life of the sensor. The oxygen sensor is automatically placed in Bypass during
conditioning, when tests are completed or when an excessively high level of oxygen is detected.

Test Completion

At some point sufficient measurements will have been made to accurately determine the transmission rate
of the samples. A Permeation Test can be terminated manually by the user, after a fixed number of test
cycles or using an automatic convergence process. When testing is complete, the cell is removed from the
test sequence and no further transmission rate examination of that sample is performed. For more
information on stopping a transmission rate test see "Determining When to Stop a Test" in Chapter 3.

Report Generation

You can request a report for one or both cells at any time until a new permeation test is started. See the
instrument Help System for information on printing reports.
MOCON, Inc.
Revision L
1-3

Advertisement

Table of Contents
loading

Table of Contents