Setting The Gas Supply Pressures; System Outgassing - Ametek mocon OX-TRAN 2/12 R Operator's Manual

Oxygen transmission rate system
Table of Contents

Advertisement

Setting Up

Setting the Gas Supply Pressures

Set the Gas Supply Pressures as follows:
1.
Set the Carrier Gas cylinder or main line regulator pressure to 35 psi.
2.
Set the Test Gas cylinder or main line regulator pressure to 35 psi.
3.
Set the pressure on the Carrier Gas regulator-tee so that the gauge reads 29 psi.
4.
Set the pressure on the Test Gas regulator-tee so that the gauge reads 29 psi.
Caution:
The maximum Carrier Gas pressure to the instrument must not exceed 32 psi.
Input pressures greater than 32 psi (2.2 bar) will damage the system.
Caution:
The maximum Test Gas pressure to the instrument must not exceed 32 psi.
Input pressures greater than 32 psi (2.2 bar) will damage the system.
Note: When using compressed gas cylinders the tank must be replaced when the pressure falls below 300
psi (on the first stage of the regulator).

System Outgassing

Before the system can be used any air trapped in the gas lines and instrument must be purged out. After
purging the carrier gas supply system, the instrument must be out-gassed. After the initial purge, residual
oxygen may still be present in the seals and void spaces within in the instrument. Out-gassing removes this
residual oxygen from the system.
To out-gas the system follow these steps:
1.
Verify the Gas Supply Pressures are set correctly; see setting the Gas Supply Pressures on page 2-
8.
2.
Turn the instrument on.
3.
Mount a film in the Test Cells, for more information see "Mounting a Sample in the Test Cell" on
page 6-5.
4.
Load and Clamp the Test Cells, for more information refer to "Loading and Unloading the Test Cell"
on page 6-6.
5.
Let the system out-gas for at least one hour before starting a test.
2-8
OX-TRAN Model 2/12 R Operator's Manual
Revision L
MOCON, Inc.

Advertisement

Table of Contents
loading

Table of Contents