OX-TRAN Model 2/12 R Operator's Manual
Instrument ReZero:
NOTE:
Use of the "Test ID", "Sample ID" and "Material ID" fields is advised to document the all of the
relevant information on the specific material used to perform the test.
7.
Verify the following Cell Level parameters have been correctly assigned:
Package/Film selection:
Area/Cell:
8.
Verify the following Test Level parameters have been correctly assigned:
Test Mode:
Exam Minutes:
Individual Zero Mode:
9.
After the film has been mounted and test parameters correctly assigned, start testing the reference
sample.
10.
Monitor the progress of the test using the "Cell Status" and "Instrument Status" screens. When the
reference sample has reached equilibrium, advance the cell to the "Test Complete" state. The sample is
usually considered to be at equilibrium when there is no discernable trend in the data.
NOTE:
Do not advance the test to the Test Complete state unless the film is at equilibrium.
MOCON, Inc.
calibrating with another known film, use the temperature at
which the transmission rate for the film was measured.
Instrument ReZero
ReZero Frequency
ReZero Exam Minutes
Verify that Package mode is selected if using a masked
MOCON Certified Film. If using another film, set the
Package/Film control to the appropriate setting for the method
used to establish the reference transmission rate.
Verify the area is set correctly when not using Package mode.
Verify the Test Mode parameter is set to continuous mode.
Set the examination time appropriately for the sample being
tested. At least 45 minutes is recommended.
Use of an "Individual Zero" test phase is not advised when
using the OX-TRAN 2/12 to test MOCON Certified Films. If an
"Individual Zero" test phase is used, setting the Individual Zero
control to "Beginning" is advised.
Revision L
Permeant Sensor Adjustment
Enabled
2
45
5-3