OX-TRAN Model 2/12 R Operator's Manual
6.
REGULATOR ASSEMBLY - If the customer chooses to purchase the optional regulator assembly, then
the two gas cylinders are the only items that must be supplied by the customer. The regulator assembly
contains everything else that is needed.
7.
Oxygen Trap - To ensure the carrier gas supplied to the OX-TRAN Model 2/12 R is free of residual
oxygen an Oxygen Trap is required for each instrument. An Oxygen Trap is included with a purchased
instrument and can be purchased directly as a standalone item (P/N 051-595) from MOCON. If the
customer would rather purchase directly from the supplier, a suitable device (P/N 202295-B) is available
from:
COPPER TUBING - Metallic tubing is required for connecting the oxygen and nitrogen supplies to the rear
8.
panel of the instrument. Desiccated and sealed (D&S), refrigeration-grade copper tubing (1/8 OD x 0.030 Wall / 3.175
mm OD x 0.762 mm) is required. This tubing can be purchased through MOCON by calling MOCON in the USA at
(763) 493-6370, or through your local tubing supplier.
DISTILLED WATER - You should use HPLC grade water for relative humidity generation. Obtain this
9.
water through a local supplier.
MOCON, Inc.
Chromatography Research Supplies Inc.
2601 Technology Drive
Louisville, KY 40299 USA
Telephone 800 327 3800
www.chromres.com
RECOMMENDED ELECTRICAL INSTALLATION
NOTES
Revision L
Site Preparation Instructions
A-3