External Current Sensor Input (/Ex1 And /Ex2 Options); D/A Output (/Da4, /Da12 Options) - YOKOGAWA WT332E User Manual

Digital power meter
Hide thumbs Also See for WT332E:
Table of Contents

Advertisement

7.4 Features
Number of displayed digits
Data update interval
Response time
Auto range monitor
Overrange display
Hold
Single update
MAX hold
7.4.7 Internal Memory
Item
Measured data
Setup information
Number of blocks which can be stored
7.5
External Current Sensor Input
(/EX1 and /EX2 options)
Item
Allows input of voltage output type current sensor signal. For detailed input specifications, see section 7.1, "Input."
7.6

D/A Output (/DA4, /DA12 Options)

Item
Output voltage
Number of output channels
Output items
Accuracy
D/A conversion resolution
Minimum load
Update interval
Temperature coefficient
7-14
Select 5 or 4 digits.
Select 0.1 s, 0.25 s, 0.5 s, 1 s, 2 s, 5 s, 10 s, 20 s or Auto.
* In the case of Auto Update Rate cannot use the Integration function and store the measurement data.
Data update rate × 2 or less
(The time it takes to reach the accuracy of the final value when the displayed value changed from 0
to 100% or 100 to 0% of the rated range)
The indicator illuminates when the input signal meets the conditions for auto range switching.
Overrange "- - - oL -" is displayed for the following conditions.
Crest factor 3, 6
• When the measured value exceeds 140% of the rated range
* WT310EH:40 A range
When the measured value exceeds 110% of the rated range
Crest factor 6A
• When the measured value exceeds two times of the above setting.
Holds the displayed value.
Updates the displayed value once each time the SINGLE key is pressed during Hold.
Holds the maximum displayed value of U, I, P, S, Q, U±pk, I±pk, and P±pk.
Specifications
Recall the stored measurement data by a communication command.
Storage interval
Data update interval or in the range of 1 s to 99 hrs 59 min 59 s.
There is no backup function for stored measurement data.
* In the case of Auto Update Rate cannot store the measurement data.
Saves/Loads four patterns of setup information.
Model
normal measurement
WT310E/WT310EH
9000
WT332E
4000
WT333E
3000
* Harmonic Measurement (/G5 Option)
Harmonic measurement function is turned ON.
Specifications
• Measurement range of the /EX1 option
Crest factor 3: 2.5 V, 5 V, 10 V
Crest factor 6 or 6A: 1.25 V, 2.5 V, 5 V
• Measurement range of the /EX2 option
Crest factor 3: 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V
Crest factor 6 or 6A: 25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V
Specifications
±5 V FS (approx. ±7.5 V maximum) against each rated value
4 outputs for products with the /DA4; 12 outputs for products with the /DA12 option
Set for each channel.
U, I, P, S, Q, λ, Φ, fU, fI, Upk, Ipk, WP, WP±, q, q±, MATH
±(accuracy of each measurement item + 0.2% of FS) (FS = 5 V)
16 bits
100 kΩ
Same as the data update interval.
* Approximately the same period as the signal's period when the data update interval is set to Auto.
But, at least 100 ms.
±0.05%/°C of FS
normal +harmonic measurement
700
300
200
IM WT310E-02EN