5.3 Producing D/A Output
Relationship between Output Items and D/A Output Voltage
Frequency
D/A output
Approx. 7.5 V
5.0 V
0.5 V
Integrated Value
D/A output
Approx. 7.0 V
5.0 V
t
: Rated integration time
0
Other Items
Displayed Value
• The range between +5 to +7 V and –5 to –7 V is not output for λ and Φ. When an error occurs, the output
is approximately ±7.5 V.
For the efficiency math function, +5 V represents 100%.
For Upk and Ipk, ±5 V represents the application of 3 times the rated range value (6 times the rated range
value when the crest factor is 6 or 6A).
When the selected output function is "----" or there is no numerical data, the output is 0 V.
5-10
0.1 Hz
10 Hz
1 Hz
100 Hz
When 140% of
the rated input is applied.
0
Output
Approx. 7.0 V
140%
5.0 V
100%
0 V
0%
–5.0 V
–100%
pprox. –7.0 V
–140%
Displayed value
100 kHz
1 kHz
10 kHz
When the rated input is applied.
Integration time
t
0
D/A output
Approx. 7.5 V
Approx. 7.0 V
5.0 V
–140
–100
0
–5.0V
Approx. –7.0 V
Approx. –7.5 V
100
140
Displayed value (%)