2.9 Wiring the Circuit under Measurement for Direct Input
5
Connect the safety terminal adapters to the WT332E/WT333E voltage input terminals.
Pay attention to the terminal polarities so that you don't attach the cables to terminals (VOLTAGE and ±)
with the wrong polarities.
Voltage input terminals
(V: VOLTAGE terminal)
The wiring for voltage measurement is complete. The wiring for measuring the power of a three-
phase, three-wire DUT is complete.
2-36
Red R-phase adapter: Connect to the VOLTAGE voltage input terminal of element 1.
Red T-phase adapter: Connect to the VOLTAGE voltage input terminal of element 3.
Black S-phase adapters: Connect to the ± voltage input terminals of elements 1 and 3.
T
S
R
Power supply
T
S
R
S
S
R
T
± V
1
3
R
T
S
R
S
T
WT332E/WT333E
1
3
SOURCE
T
R
S
T
DUT
R
S
T
C
±
I1
V
R
U1
±
S
±
U3
V
I3
C
±
LOAD