7.4

Features

7.4.1 Voltage, Current, and Active Power Measurements
Item
Measurement method
Crest factor
Wiring system
Range select
Auto range
Display mode switching
Measurement
synchronization source
Line filter
Peak measurement
Zero-level compensation
IM WT310E-02EN
Specifications
Digital sampling method
3 or 6 (6A)
WT310E, WT310EH (One element model)
Single-phase, two-wire (1P2W)
WT332E (Two element model)
Select single-phase, two-wire (1P2W); single-phase, three-wire (1P3W); or three-phase, three-
wire (3P3W).
WT333E (Three element model)
Select single-phase, two-wire (1P2W); single-phase, three-wire (1P3W); three-phase, three-wire
(3P3W); three-phase, four-wire (3P4W); or three-voltage, three-current (3V3A).
Select manual or auto ranging.
Range increase
The range is increased when any of the following conditions is met.
• Crest factor 3
Urms or Irms exceeds 130% of the currently set measurement range.
Upk , Ipk value of the input signal exceeds 300% of the currently set measurement range.
• Crest factor 6
Urms or Irms exceeds 130% of the currently set measurement range.
Upk , Ipk value of the input signal exceeds 600% of the currently set measurement range.
• Crest factor 6A
Urms or Irms exceeds 260% of the currently set measurement range.
Upk , Ipk value of the input signal exceeds 600% of the currently set measurement range.
On the WT332E/WT333E, when any of those input elements meets the above condition, the
range is increased the next time the measured value is updated.
Range decrease
The range is decreased when all of the following conditions are met.
• Crest factor 3
Urms or Irms is less than or equal to 30% of the measurement range.
Urms or Irms is less than or equal to 125% of the next lower measurement range.
Upk , Ipk value of the input signal exceeds 300% of the currently set measurement range.
• Crest factor 6 or 6A
Urms or Irms is less than or equal to 30% of the measurement range.
Urms or Irms is less than or equal to 125% of the next lower measurement range.
Upk , Ipk value of the input signal exceeds 600% of the currently set measurement range.
On the WT332E/WT333E, when all of the input elements meet the above conditions, the range
is decreased the next time the measured value is updated.
Select RMS (the true RMS value of voltage and current), VOLTAGE MEAN(the rectified mean value
calibrated to the RMS value of the voltage and the true RMS value of the current), or DC (simple
average of voltage and current).
Select voltage, current, or the entire period of the data update interval for the signal used to achieve
synchronization during measurement.
In the case of Auto Update Rate, select the voltage or current from the equipped element.
Select OFF or ON (cutoff frequency at 500 Hz).
Measures the peak (max, min) value of voltage, current, or power from the instantaneous voltage,
instantaneous current, or instantaneous power that is sampled.
Removes the internal offset of this instrument.
1
2
3
4
5
6
7
App
Index
7-9