Probe Compensation

When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or errors. The
probe compensation procedures are as follows:
1.
Perform Step 1, 2, 3 and 4 specified in "Function Inspection".
2.
Check the displayed waveforms and compare them with Figure 1-10.
Over compensated
3.
Use a nonmetallic screwdriver to adjust the low-frequency compensation
adjustment hole on the probe until the waveform is displayed as "Perfectly
compensated" in the figure above.
MSO5000 User Guide
Perfectly compensated
Figure 1-10 Probe Compensation
Under compensated
RIGOL
1-9