Single-channel mode: only one of the four channels (CH1/CH2/CH3/CH4) is
enabled.
Half-channel mode: either CH1 or CH2 is enabled; and either CH3 or CH4 is
enabled.
All-channel mode: CH1/CH2 are both enabled or CH3/CH4 are both enabled.
Note that the currently enabled channel should be used as the trigger source at
this time.
Note:
The sample rate is displayed both in the status bar at the upper part of the
screen and in the CH SampleRate menu. To indirectly modify the sample
rate, rotate the Horizontal
base or modify the memory depth.
The impact of low sample rate on the waveform:
1. Waveform Distortion: when the sample rate is too low, some waveform
details are lost, and the sample waveform displayed is rather different from the
actual waveform of the signal.
2. Waveform Aliasing: when the sample rate is twice lower than the actual
signal frequency (Nyquist Frequency), the frequency of the waveform rebuilt from
the sample data is smaller than the actual signal frequency.
3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt
from the sample data does not reflect all the actual signal information.
4-8
Chapter 4 To Set the Sample System
SCALE knob to adjust the horizontal time