The following equation shows the relations among memory depth, sample rate, and
horizontal time base scale:
——indicates the memory depth. The unit is pts.
MDepth
——indicates the sample rate. The unit is Sa/s.
SRate
——indicates the horizontal time base scale. The unit is s/div.
TScale
——indicates the number of grids in the horizontal direction. The unit is div.
HDivs
Therefore, under the same horizontal time base scale, a higher memory depth can
ensure a higher sample rate.
Press Acquire → Mem Depth, then rotate the multifunction knob
the desired memory depth, and then press down the knob to select the option. You
can also press the Mem Depth key continuously or enable the touch screen to
select the memory depth. By default, the memory depth is Auto. The memory depth
of the analog channel is related to the current channel mode.
In single-channel mode, the memory depths available include Auto, 1 Kpts, 10
Kpts, 100 Kpts, 1 Mpts, 10 Mpts, 25 Mpts, 50 Mpts, 100 Mpts and 200 Mpts.
In half-channel mode, the memory depths available include Auto, 1 Kpts, 10
Kpts, 100 Kpts, 1 Mpts, 10 Mpts, 25 Mpts, 50 Mpts and 100 Mpts.
In all-channel mode, the memory depths available include Auto, 1 Kpts, 10 Kpts,
100 Kpts, 1 Mpts, 10 Mpts, 25 Mpts and 50 Mpts.
Note:
In "Auto" mode, the oscilloscope selects the memory depth automatically
according to the current sample rate.

LA Memory Depth

Press Acquire → More to view the memory depth of the current digital channel
under LA Mem Depth. The LA memory depth will change with the memory depth of
4-10
Trigger Point
T
Pre-sample
Memory Depth
Figure 4-4 Memory Depth
=
MDepth SRate TScale HDivs
Chapter 4 To Set the Sample System
Delayed Sample
to switch to
MSO5000 User Guide