Chapter 5 To Trigger the Oscilloscope
Trigger Condition:
Press When and rotate the multifunction knob
Setup/Hold trigger. Press down the knob to select the trigger condition. You can also
press When continuously or enable the touch screen to select it.
Setup: the oscilloscope triggers when the setup time is smaller than the
⚫
specified setup time. After selecting this type, press Setup and rotate the
multifunction knob
numeric keypad to set it.
Hold: the oscilloscope triggers when the hold time is smaller than the specified
⚫
hold time. After selecting this type, press Hold and rotate the multifunction
knob
to set the hold time. You can also use the pop-up numeric keypad to
set it.
Setup/Hold: the oscilloscope triggers when the setup time or the hold time is
⚫
smaller than the specified time value. After selecting this type, press Setup and
Hold and rotate the multifunction knob
respectively. You can also use the pop-up numeric keypad to set them.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch
the current trigger mode. For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameter (noise rejection) under this trigger type. For details, refer
to descriptions in "Noise Rejection".
Trigger LEVEL/Threshold Level:
Rotate the Trigger
to "Trigger LEVEL/Threshold Level". The current trigger level/threshold level
value is displayed at the upper-right corner of the screen.
MSO5000 User Guide
to set the setup time. You can also use the pop-up
LEVEL knob to adjust the trigger level/threshold level. Refer
to select the trigger condition of
to set the setup time and hold time
RIGOL
5-31