also press Trig Source continuously or enable the touch screen to select it.
The available trigger sources include Internal and Manual. For the Infinite
burst type, the Trig Source menu is disabled and grayed out.
Internal: the Function/Arbitrary Waveform Generator outputs
continuous N cycle bursts.
Manual: when you press Manual Trig for one time, one burst output is
immediately launched for the corresponding channel. For the Infinite
burst type, the Cycles menu is disabled and grayed out.
Set the burst period
It is defined as the time from the beginning of one burst to that of another
burst.
Press Burst Period, then rotate the multifunction knob
pop-up numeric keypad to set the desired burst period. Its available range
is from 1 us to 500 s. By default, it is 10 ms.
17-20
Chapter 17 Function/Arbitrary Waveform Generator (Option)
or use the
MSO5000 User Guide