Burst - Rigol MSO5072 User Manual

Digital oscilloscope
Hide thumbs Also See for MSO5072:
Table of Contents

Advertisement

Chapter 17 Function/Arbitrary Waveform Generator (Option)

Burst

The built-in Function/Arbitrary Waveform Generator of MSO5000 series oscilloscope
supports outputting the burst waveforms with a specified number of cycles. It
supports using internal and manual trigger sources to control the output of the burst
waveforms. Two burst types are supported: N Cycle and Infinite.
Press Settings → Type, and rotate the multifunction knob
and then press down the knob to select it. You can also press Type continuously or
enable the touch screen to select it.
1. Set the Burst Type
Press Burst Type, and rotate the multifunction knob
burst type, and then press down the knob to select it. You can also press Burst
Type continuously or enable the touch screen to select it. The available burst
types include N Cycle and Infinite.
N Cycle: outputs the burst waveforms with a specified number of cycles
once receiving the trigger signal.
Infinite: sets the number of cycles to Infinite. It outputs continuous
waveforms once receiving the trigger signal.
2. Set the Number of Cycles
When the burst type is set to "N Cycle", you need to set the number of cycles.
Press Cycles, then rotate the multifunction knob
keypad to set the number of cycles. For the Infinite burst type, the Cycles menu
is disabled and grayed out.
3. Set the Burst Delay
Burst delay indicates the time from when receiving the trigger signal to starting
to output N Cycle of bursts or Infinite burst.
Press Delay, then rotate the multifunction knob
keypad to set the desired delay time. The range of the delay time related to the
burst cycles.
4. Burst Setting
Press Settings to enter the burst setting menu.
Set the trigger source
The trigger sources of the burst can be Internal or Manual. When the
Function/Arbitrary Waveform Generator receives a trigger signal, a burst
output is generated, and then it waits for another trigger signal.
Press Trig Source and rotate the multifunction knob
desired trigger source, and then press down the knob to select it. You can
MSO5000 User Guide
to select "Burst",
to select the desired
or use the pop-up numeric
or use the pop-up numeric
to select the
RIGOL