To Save To Internal Memory; To Clear The Display Of The Reference Waveform; To View Details Of The Reference Waveform - Rigol MSO5072 User Manual

Digital oscilloscope
Hide thumbs Also See for MSO5072:
Table of Contents

Advertisement

Chapter 12 Reference Waveform
Press VOffset, then rotate the multifunction knob
numeric keypad to set the vertical offset of the reference waveform. You can
also enable the touch screen and use the "Drag" gesture to adjust the vertical
offset.

To Save to Internal Memory

Press SaveToRef to save the displayed waveform for the specified source to the
internal memory as the reference waveform.
Note:
This operation only saves the reference waveform to the volatile memory, and
the waveform will be cleared at power-off.

To Clear the Display of the Reference Waveform

Press Clear to clear the display of the current reference waveform on the screen.
Meanwhile, the VScale and VOffset menus for the reference waveform are grayed
out and disabled. You can also press CLEAR on the front panel to clear the display of
the reference waveforms of all the reference channels.
Note:
This operation only clears the display of the reference waveform on the screen,
and the reference waveform is still saved in the memory.

To View Details of the Reference Waveform

Press More → Details, and the detailed information about all the reference
waveforms will be displayed on the screen in the form of a list. Press Details again
to close the details list. You can also enable the touch screen to tap the icon
the right-upper corner of the table to close it.
MSO5000 User Guide
RIGOL
or use the pop-up
at
12-3