7.12.5 Pump-motor supervision: Overload or blockage


Pump alarm
bL0C
Command

7.12.6 Pump-motor supervision: Dry running


Pump alarm
PuLEU
Command
03/2022
 The SelfCheck Assistant monitors the Varioflex pump:
1. Alarm sounds as dual-tone signal for pump-motor overload or blockage.
2. Display of
°C
3. The red LED
 Heater switches off on both poles,
 Pump and refrigerating unit are switched off electronically.
 Find the cause of the fault. Perhaps the viscosity of the heat transfer liquid is
too high or the pump is blocked.
 Press the Enter key.
 Also press this key if the unit has been switched off in the fault state.
Pump-motor alarm!
Pump-motor alarm is shown in the display and signifies that unlocking is
only possible on the Master control panel.
 The SelfCheck Assistant monitors the Varioflex pump:
1. Alarm sounds as dual-tone signal when the pump runs without liquid. This
can only occur when the float level measurement has failed.
2. The display of
detected a pump low level.
°C
3. The red LED
 Heater switches off on both poles,
 Pump and refrigerating unit are switched off electronically.
 The cause of the failure of the level measurement with the floatation sensor must be found and
rectified. Perhaps foreign bodies in the bath block it.
 Press the Enter key.
 Also press this key if the unit has been switched off in the fault state.
Alarm! Low level (pump)
Alarm! Low level (pump) is shown in the display and signifies that unlocking
is only possible on the Master control panel.
bL0C
signals blockage.
above the fault triangle
PuLEU
signals that the SelfCheck Assistant has
above the fault triangle
Proline Kryomats
flashes.
flashes.
93 / 137