Lauda PROLINE Kryomats RP 4050 C Operation Manual page 103

Hide thumbs Also See for PROLINE Kryomats RP 4050 C:
Table of Contents

Advertisement

OUT_MODE_00_X
OUT_MODE_01_X
OUT_MODE_03_X
OUT_MODE_04_X
START
STOP
RMP_SELECT_X
RMP_START
RMP_PAUSE
RMP_CONT
RMP_STOP
RMP_RESET
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_X
RMP_OUT_02_XXX
Permitted data formats:
-XXX.XX
-XX.XX
-X.XX
-.XX
03/2022
Command
 For "_" use also " " (blank character).
 Response from thermostat "OK" or in case of error " ERR_X" (RS 485 interface e.g.
"A015_OK" or in case of error "A015_ERR_X".).
 The command from the computer must be terminated with CR, CRLF, or LFCR.
 The response of the thermostat is always terminated with CRLF.
 After each command sent to the thermostat, it is necessary to wait for the reply before
sending another command. This ensures that the sequencing of inquiries and answers is
clear.
CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
-XXX.X
-XXX.
-XX.X
-XX.
-X.X
-X.
-.X
.XX
Keyboard Master: 0 = free / 1 = locked (corresponds to "KEY").
Control: 0 = internal / 1 = external Pt100 / 2 = external analogue / 3 = external serial / 5 =
external Ethernet / 6 = external EtherCAT.
Note: With some temperature control devices this command can only be carried out if the
command OUT_PV_05_XXX.XX has been sent by the interface. Also notice that the
command OUT_PV_05_XXX.XX must be cyclically transmitted by the selected interface.
Keyboard Command remote control: 0 = free / 1 = locked.
Setpoint offset source: 0 = normal / 1 = external Pt / 2 = external analogue / 3 = external
serial.
Note: With some temperature control devices this command can only be carried out if the
command OUT_PV_05_XXX.XX has been sent by the interface. Also notice that the
command OUT_PV_05_XXX.XX must be cyclically transmitted by the selected interface.
Switches the unit on (after Standby). See safety information ( 7.5.3).
Switches the device into Standby (pump, heater, cooling unit OFF).
Selection of the program (1 – 5) to which the further instructions apply. When the unit is
switched on, program 5 is selected automatically.
Start the programmer.
Hold (pause) the programmer.
Restart the programmer after pause.
Terminate the program.
Delete the program (all Segments).
Set a programmer segment (temperature, time, tolerance and pump level). A segment is
added and appropriate values are applied to it.
Number of times the program runs: 0 = unlimited / 1 – 250.
-XXX
XXX.XX
-XX
XX.XX
-X
X.XX
.X
Proline Kryomats
Explanation
XXX.X
XXX.
XX.X
XX.
X.X
X.
XXX
XX
X
103 / 137

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents