Lauda PROLINE Kryomats RP 4050 C Operation Manual page 68

Hide thumbs Also See for PROLINE Kryomats RP 4050 C:
Table of Contents

Advertisement

Edited program example (see dashed curve in the graph on previous page).
No
T end °C
Start
30.00°C
1
30.00°C
2
50.00°C
3
50.00°C
4
70.00°C
5
70.00°C
6
60.00°C
7
30.00°C
Pump
Menu
 Insert new segment ( Section 0)
  Change segment time or tolerance ( Section 0)
The field tolerance (refer to the above program table and the graph below):
 It facilitates exact conformance to the dwell time at a specified temperature. Segment 1 is not
 A tolerance range which is too tight can however also cause undesired delays. In particular with
 Only flat (slow) ramps should be programmed where necessary with a tolerance range. Steep
°C
80
70
60
50
40
30
20
 
10
0
20
40
68 / 137
Each program begins with the segment "Start". It defines at which temperature Segment 1 is to
continue the program. It is not possible to specify a time for the Start segment. For thermostats
without cooling ability, the start temperature must be selected higher than the bath temperature,
which prevails before the program start. Without the Start segment, Segment 1 would be
different depending on the bath temperature at the start of the program.
Time [h:m]
Tolerance
-------
0.00°C
00:20
0.10°C
0.00°C 
00:20
0.10°C 
00:20
00:20
0.00°C
00:10
0.80°C
00:30
0.00°C
00:00
0.00°C
End
Insert
Delete
processed until the bath temperature is within the tolerance range , so that the ramp (Segment
2) starts delayed at .
external control the range should not be chosen too tightly. In Segment 5 a larger tolerance has
been entered, so that the desired time of ten minutes is maintained even with settling action .
ramps which lie close to the maximum possible heating or cooling rates of the thermostat may be
severely delayed by a tolerance range that is too tight (here in Segment 2) .
60
80
100
120
Proline Kryomats
No
Pump
Start
-------
1
2
3
4
5
6
7
Pump
Example for the influence of the tolerance field input in
case of external bath temperature control:
The setpoint temperature of the programmer is shown in
grey.
The actual temperature in the external bath container is
represented as a continuous line.
t/min
Out 1
Out 2
-------
-------
2
-------
-------
2
-------
-------
2
-------
-------
2
-------
-------
2
-------
-------
2
-------
-------
2
-------
-------
Menu
End
Insert
Out 3
-------
-------
-------
-------
-------
-------
-------
-------
Delete
03/2022

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents