No.
Pump
Start
-------
1
Pump
Menu
End of segment temperature:
Min: -150.00°C Max:450.00°C
1
Help
6
Menu
Input segment time:
003:00
Hours(max.999):Minutes
1
Help
6
Menu
03/2022
Out 1
Out 2
-------
-------
4
-------
-------
End
Insert
25,
00
2
3
4
7
Screen
8
T
9
set
2
3
4
7
Screen
8
T
9
set
Then continue with
Out 3
-------
The right-hand part of the entry table appears
-------
Here, in the "Pump" field, the pump level and,
Delete
A new segment is produced by moving the cell
If the field in the column
5
T
0
fix
Enter the value, confirm with the
If the field in the column
If 0 is entered into the field "Time", "------"
Enter the segment time and confirm with the
5
T
0
fix
Continue to the "Tolerance" entry field with
Proline Kryomats
to the pump and signal
output setting.
as shown on the left.
in the fields "Out 1" to "Out 3", the contact
outputs of the contact mode (accessory) can
be programmed. With the setting "-------"
the starting value is retained which was either
set before the program start or was defined by
a previous segment in the running program.
Further details are given on the following pages.
with the black background to a blank line with
the cursor keys and then pressing the soft key
Insert . The values of the cell located above
it are automatically copied.
T end °C
background, the entry mode "End of segment
temperature" is obtained by pressing the
key. Depending on the setting, that is the
temperature, which the thermostat is to
achieve on the internal or external temperature
probe.
continue to the "Time" entry field with
Time
background, the entry mode for the "Segment
time" time setting is obtained by pressing the
key.
appears. Then the final temperature is
approached as quickly as possible. With a time
entry the final temperature is obtained exactly
after the time expires (ramp).
key.
.
has a black
key and
.
" has a black
73 / 137