LAUDA Thermostats largely require no service. If the heat transfer liquid becomes contaminated, it should be
replaced ( 6.2).
03/2022
Uncontrolled start-up of the pump when jamming released
Crushing, electric shock hazard
Disconnect the device from the mains power supply before
the repair.
Have the repair carried out only by a specialist.
Contact with hot or cold device parts, heat transfer liquid or
Allow device parts, accessories and heat transfer liquid to
reach room temperature before touching them.
Have repairs carried out only by a specialist.
At the back of the Master head a main fuse switch
mains connection when an overload occurs. It is then in the "O" position and can be set in
the "–" position again.
If the fuse trips again, Service must locate the cause.
Additionally, a safety fuse, which protects the low voltages, is situated on the mains board. If
a fuse fails ( mains lamp does not light) only replace with a fuse with the specified data
(one x T (= slow-blow) 10 A, size 5 x 20 the Fuse is located in the unit as shown below).