Lauda PROLINE Kryomats RP 4050 C Operation Manual page 105

Hide thumbs Also See for PROLINE Kryomats RP 4050 C:
Table of Contents

Advertisement

VERSION_B
VERSION_T
VERSION_A
VERSION_V
VERSION_D
VERSION_M_0
VERSION_M_1
VERSION_M_3
VERSION_M_4
VERSION_M_5
STATUS
STAT
RMP_IN_00_XXX
RMP_IN_01
RMP_IN_02
RMP_IN_03
RMP_IN_04
RMP_IN_05
LOG_IN_00_XXXX
LOG_IN_01
LOG_IN_02
LOG_IN_03
03/2022
Command
Read software type of Command.
Read software type of cooling system.
Read software type of analogue module.
Read software type of RS 232/485 module.
Read software type of digital module.
Read software type of solenoid valve (cooling water).
Read software type of solenoid valve (automatic refill).
Read software type of solenoid valve (shut-off valve 1)
Read software type of solenoid valve (shut off valve 2).
Read software type of high temperature cooler.
Read equipment status 0 = OK, -1 = error.
Read error diagnosis response:
XXXXXXX  X = 0 no error, X = 1 error.
1. char = error.
2. char = Alarm.
3. char = Warning.
4. char = over temperature.
5. char = low-level error.
6. char = high-level error (at adjustment alarm).
7. char = no external control variable.
Read a program segment XXX
(response: e.g. 030.00_010.00_005.00_001.00  set point temperature 30.00 °C,
time = 10 min, tolerance = 5.00 °C, pump level = 1).
Read the current segment number.
Read the set number of program runs.
Read the current program run.
Read the program to which further instructions apply.
Read which program is currently running (0 = none).
Query a measuring point XXXX from data logger
(Reply: e.g. 020.00_021.23_030.50 => set point temperature = 20.00 °C, bath
temperature = 21.23 °C, external temperature = 30.5 °C).
Read all measuring points from data logger
As a difference to the command "LOG_IN_00", a tabulator is used here as separator instead
of ,_' . The measuring points are separated by CR and LF. The end is marked by CR LF CR
LF.
Read the start time from data logger
(Reply: e.g. 20_14_12_20  day 20, time 14:12:20).
Read the acquisition interval from the data logger (Reply in seconds).