Lauda PROLINE Kryomats RP 4050 C Operation Manual page 34

Hide thumbs Also See for PROLINE Kryomats RP 4050 C:
Table of Contents

Advertisement

Ammonia (NH
)
3
Iron (Fe), dissolved
Manganese (Mn), dissolved
Aluminum (Al), dissolved
Free aggressive carbonic acid (CO
Hydrogen sulfide (H
Algae growth
Suspended matter
Risk to the environment due to oil contamination of the cooling water circuit
With a leaky condenser there is the danger that refrigerating machine oil from the coolant circuit of the cooling
thermostat can pass into the cooling water.
Follow all the legal requirements and the regulations of the water supply utility, which apply at the point of use.
Water pollution due to leakage
To avoid pollution due to a leak in the cooling water system it is recommended that a leakage-water detector
with a water cut-off is installed.
Servicing intervals
Follow the information for cleaning and decalcifying the cooling water circuit ( 9.3.4.2).
c) Hoses
Elastomer hoses
Hose type
EPDM hose
uninsulated
EPDM hose
insulated
Silicone hose
uninsulated
Silicone hose
insulated
34 / 137
)
2
S)
2
Internal width Ø mm
12
12
External Ø.
approx. 35mm
11
11
External Ø.
approx. 35mm
 EPDM hose is not suitable for mineral oils.
 With silicone rubber, silicone oils lead to substantial swelling  never use silicone oil with
silicone hoses.
 Secure hoses against slippage with hose clips.
Not permissible
< 0.2 mg/L
< 0.05 mg/L
< 0.2 mg/L
Not permissible
Not permissible
Not permissible
Not permissible
Temperature range °C
10 – 90
-35 – 90
10 – 100
-60 – 100
Proline Kryomats
Field of application
For all heat transfer liquids
except mineral oils
Water,
Glycol/water mixture
Order number
RKJ 112
LZS 021
RKJ 059
LZS 007
03/2022

Hide quick links:

Advertisement

Table of Contents
loading