9.1.5 Fault memory

For the analysis and localization of faults the Command version includes a fault memory in which up to 45 fault and
alarm messages are saved.
Command
No. Source Code Type Date
10 Safety 2 Alarm ------------ -----------
9 Safety 4 Warn. 28.08.03 15:32:02
8 Contro. 32 Error
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Pump
Menu
9.2

Cleaning

9.2.1 Cleaning the surface of the device

Warning!
Notice
Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been added and
using a damp cloth.
03/2022
17.07.0310.:52:02
End
T
set
Live parts in contact with cleaning agents
Disconnect the device from the mains before cleaning.
Live parts in contact with cleaning agents
Disconnect the device from the mains before cleaning.
Water and other liquids must not enter the control head.
Carry out appropriate decontamination if hazardous material is spilt on or in the equipment.
The cleaning or decontamination method is determined by the user's specialist knowledge. In case of
doubt, contact the manufacturer.
Time
store
 The last message is at the top.
 Each message line can be marked with the
 Under "Source", the CAN node is displayed
 Code is the number, which in the Master is
T
fix
 Type: Alarm, Warning or Fault (Error).
Electric shock hazard
Property damage
Proline Kryomats
Error store
Settings
Device status
Display
.
cursor keys. The message appears in plain text
in the footer.
which signaled the fault.
shown in the display until the cause has been
rectified.
Error
111 / 137