5.4 Step 4: Software Update

20
Step 4: Software Update
 If the "No Software Available"
pop window is displayed:
 The connected Kendall SCD
700 series controller has the
latest controller software.
 No new controller software is
available.
 Select "Okay" to return to the
Home Screen.
.
Step 4: Software Update
(continued)
 The "Select Software Version"
field will be automatically
populated for users who are not
Covidien employees.
 On the Kendall SCD 700 series
controller, simultaneously
hold down A, B, & Right
Arrow buttons.
Step 4 (continued): Software
Update
 Press the power button (once)
while still holding down the A,
B, & Right Arrow buttons.
 Release the power button while
continuing to hold down the A,
B, & Right Arrow buttons for 5
seconds.