Kendall Scd™ 700 Controller; Visual Inspection - Cardinal Health Kendall SCD 700 Manual

Table of Contents

Advertisement

Section 2 – Instructions for Customer Returned Units (continued)
KENDALL SCD™ 700 SYSTEM CONTROLLER

VISUAL INSPECTION

Inspect the DME supplier contact information label. If the DME supplier contact information
label is torn, ripped or illegible, replace with a new label (see sketch below):
Inspect the power cord for damage. If power cord is frayed or damaged, send the controller
to the Covidien service center for repair (Reference Attachment A).
Inspect the controller for marks, scuffs and scratches especially on the control panel (see
below).
Minor surface scratches or scuffs are acceptable. If any marks are present that obscure or
interfere with the display of the icons or the clear membrane on the control panel, send the
controller to the Covidien service center for repair (Attachment A).
Inspect the controller for missing, broken or cracked components. If any components are
missing, broken or cracked, send the controller to the Covidien service center for repair
(Attachment A).
Inspect the controller for the presence of the serial label on the bottom of the device. It the
serial label is not present, or if it is illegible or damaged, send the controller to the Covidien
service center for repair (Attachment A).
PLACE LABEL WITHIN THE BOXED AREA
CONTROL PANEL