Section 2 – Instructions for Customer Returned Units (continued)
PREPARATION (continued)
TUBING SET LEAK TEST (continued):
Turn ball valve to open position, so that the knob position is in-line with valve
body (see picture below).
Adjust the input air pressure so that a pressure of between 125mmHg (2.4 psi)
and 135mmHg (2.6 psi) is indicated on the fixture's pressure gauge, and allow to
hold for approximately 2 seconds.
Close ball valve, so that the knob position is perpendicular to valve body (see
picture below). Activate the stopwatch.
Wait 15 seconds and observe the pressure gauge. Because the test was
conducted with known good tubing set, the pressure should not fall below
89mmHg (1.72 psi). If pressure falls below 89mmHg, contact your sales rep for
replacement tubing set Leak Test Fixture.