Section 2 – Instructions for Customer Returned Units

REFERENCE DOCUMENTS

Kendall SCD™ 700 system Operation and Service Manual

EQUIPMENT LIST

Regulated air source (i.e. compressor or factory air)
Tubing set Leak Test Fixture (provided by Covidien)
Known good tubing set

PREPARATION

TUBING SET LEAK TEST FIXTURE SETUP:
NOTE: This is to be performed upon receipt of the fixture, to confirm acceptable fixture
functionality.
Confirm the tubing set Leak Test Fixture's certification from the fixture manufacturer.
o
Confirm pressure gauge is calibrated. If not, send out for re-calibration.
o
Conduct the leak test with the known good tubing set as follows:
o
Ensure that the Leak Test Fixture (see sketch below) is connected to a regulated
air source. The input air should be at pressure between 3 psi and 4 psi.
Connect the tubing set to the connectors on the test fixture. See pictures below.