Cardinal Health™ SVED®
Negative Pressure Wound Therapy
Clinician Quick
Reference Guide
Battery Charging Port
Plugged In
Power/Charging Status
Low Pressure/Leak Alert
Therapy Timer/Display
NOTE: Please refer to the Cardinal Health™ SVED® CLINICIAN User Manual for
detailed instructions and review the Precautions and Contraindications before use.
Insert Canister
Insert canister in holder by sliding it up into bottom of the Canister Holder. Line up
the two short ports with the Canister Holder and press until it "clicks" into place and
locks. To remove canister, press Canister Release Button, grasp canister at bottom
and pull downward.
Connect the Dressing
Connect the blue end of the tubing to the open port of the canister. Gently twist and
push connector on just enough to seal it. Make sure that the clamp on the tubing is
open.
NOTE: Please refer to the Cardinal Health™ SVED® Clinician User Manual for
instructions on how to properly dress the wound and troubleshoot issues.
Turning the SVED® ON
Press the ON
Button to start therapy. Select desired negative pressure wound
therapy setting by pressing the corresponding Pressure Setting -70, -120, or
-150mmHg
button. The Pressure Setting Button flashes until desired
pressure is reached, after which the button remains lit.
For questions or additional information, contact Cardinal Health at 1.866.484.6798
Therapy Lockout
To lock therapy setting, press and hold the desired Pressure Setting
seconds. The SVED® beeps three times indicating therapeutic setting is locked. Repeat
process to unlock therapy setting.
Intermittent Mode
To operate the SVED® in Intermittent Mode, press and hold the desired Pressure Setting
Button then momentarily press the OFF
the SVED® beeps twice indicating the SVED® is now operating in Intermittent Mode. Release
both buttons. Repeat process to return to Continuous Mode.
Troubleshooting Alerts
If the SVED® begins to alert, please refer to troubleshooting table on the back.
Turning the SVED® OFF
To turn the SVED® off, press the OFF
Button.
Notice to Users: As with any prescription medical device, failure
to follow product instructions or adjusting settings and performing
therapy applications without the express direction and/or
supervision of your trained clinical caregiver may lead to improper
product performance and the potential for serious or fatal injury.
Defibrillation: Remove the wound dressing if defibrillation is
required in the area of dressing placement. Failure to remove the
dressing may inhibit electrical current transmission and/or patient
resuscitation.
Canister Release Button
ON Button
OFF Button
Canister Full Alert
Canister Holder
Pressure Settings
Button for 3
Button. The ON
Button begins flashing and
Magnetic Resonance Imaging (MRI): The SVED® is not
MRI-compatible. Do not take into the MRI area.
Hyperbaric Oxygen Therapy (HBO): NEVER allow an SVED® —
whether on or off — inside a hyperbaric chamber. The SVED®
must be disconnected from the patient prior to HBO treatment.

Advertisement

loading

Summary of Contents for Cardinal Health SVED 6701132

  • Page 1 Therapy Timer/Display Pressure Settings Therapy Lockout NOTE: Please refer to the Cardinal Health™ SVED® CLINICIAN User Manual for detailed instructions and review the Precautions and Contraindications before use. To lock therapy setting, press and hold the desired Pressure Setting Button for 3 seconds.