DX-D 100 Mobile X-ray Unit
User Manual
2.7.8

DESIGNATED SIGNIFICANT ZONES OF OCCUPANCY

Illustration 2-1
Radiographic Examination on the Chest Unit or Front Panel
S
S = SIGNIFICANT ZONE OF OCCUPANCY
MINIMUM AREA 60 x 60 cm
MINIMUM HEIGHT ABOVE THE FLOOR 200 cm
S
Dosimeter
d
d = DISTANCE FROM THE AXIS OF THE X-RAY
BEAM TO THE DOSIMETER
SIGNIFICANT ZONE OF OCCUPANCY
AT THE LEFT SIDE OF THE CHEST UNIT
22
X-Ray equipment specified for any radiological examination that requires the
operator or staff to be close to the patient during normal use (a.e. some pediatric
examinations or other types of examinations for patients that may require
assistance), shall have at least one "Significant Zone of Occupancy" for the use
of the operator and staff, designated as follows:
CHEST UNIT
Focal Spot
Focal Spot
CHEST UNIT
S1
Focal Spot
MOBILE X- -RAY UNIT
Protective Device
Wall
Patient Support
X--Ray Receptor
S2
Phantom
AT THE RIGHT SIDE OF THE CHEST UNIT
MOBILE X- -RAY UNIT
SIGNIFICANT ZONE OF OCCUPANCY
0188K EN 20201116