DX-D 100 Mobile X-ray Unit
User Manual
Generator Model
Power Line Operation
Maximum Input Power
Operation independent from
mains supply (Stand-Alone)
Battery Capacity
for the Generator
Battery Capacity
for the Motors
Radiation Output Accuracy
(Reproducibility related to loading factors)
Maximum Symmetrical
Radiation Field
Maximum Heat Output
Storage / Transport
Environmental Conditions
Operating
Environmental Conditions
78
100 / 110 / 120 / 127 / 220 / 230 / 240 V~ -- Single-Phase 50 / 60 Hz
Automatic Line Compensation
Connection to standard outlets with GND that complies with local regulations
The General Circuit Breaker installed in the Mobile Unit is 10 A (1P+N curve type D),
the Power Line Installation should be provided with a Differential of 30 mA Sensitivity
and with a Thermomagnetic Switch / Circuit Breaker of:
² 13 A (curve type D) or ² 20 A (curve type C) or ² 32 A (curve type B)
Power Line Impedance must be less than the maximum indicated value:
Batteries fully charged float voltage of approx. 420 Volts at nominal of approx. 382 Volts.
Charge Capacity is: 14 Ah for Lead-Crystal Batteries
The required time for the Batteries to be fully charged is approximately:
The maximum Storage Energy Capacity is: 137500 mAs @ 80 kVp
(This is the maximum energy available for making Exposures and supplying energy to the Generator)
The Mobile Unit in Stand-Alone (disconnected from the mains) will be 100% discharged from full charge
in approximately: 9 hours for Lead-Crystal Batteries
Batteries fully charged float voltage of approx. 112 Volts at nominal of approx. 102 Volts.
The required time for the Batteries to be fully charged is 6 hours.
With the Batteries fully charged and disconnected from the mains,
the Mobile Unit can be in continuous movement during 4 hours (around 20 km).
If the Mobile Unit is left on in Stand-Alone (disconnected from the mains)
during 40 hours, it will be 100% discharged from full charge.
Measured at 75 kV: 200 mm in "X" axis and 260 mm in "Y" axis.
Measured at 125 kV: 200 mm in "X" axis and 260 mm in "Y" axis.
(Test performed at a distance from the Focal Spot of 1200 mm, in accordance with IEC 60806:1984).
Relative Humidity range of 20% to 90%
Atmospheric Pressure range of 700 hPa to 1060 hPa
(the recommended temperature for a longer life cycle of batteries is:
o
o
15
C
25
C for Lead-Crystal Batteries and 22
~
Relative Humidity (no condensing) range of 30% to 75%
Atmospheric Pressure range of 700 hPa to 1060 hPa
DX-D 100
10% V~
¦
1.2  for 110 V~, 2.5  for 230 V~
1.5 kVA
Standard
9 hours for Lead-Crystal Batteries
Charge Capacity 9 Ah
C.V. (Coefficient of variation)
260 W (1130 BTU/h)
o
Temperature range of --15
C to 40
o
Temperature range of 10
C to 35
o
C
0.05
o
C
o
C
o
25
C for Gel Batteries)
~
0188K EN 20201116