DX-D 100 Mobile X-ray Unit
User Manual
3.2.2
EMERGENCY STOP
.
Note
3.2.3
POWER LINE CONNECTION LAMP
3.2.4
COLLIMATOR LAMP
44
In the event of an emergency, the Unit is turned OFF by forcibly pressing this
switch (red mushroom-shaped switch).
The Emergency Stop must not be used to switch OFF the Unit to avoid
damaging the software. The switch is protected by a safety shield in order to
prevent it from being accidentally pressed.
For moving the Unit or charging the batteries, this device
should not be pressed.
It indicates that the Mobile Unit is connected to the mains power supply for
battery charging whenever the Line Circuit Breaker for Charging Circuits is in
the "ON" position and the Emergency Switch-Off is not pressed.
IF THIS INDICATOR IS OFF DURING THE BATTERIES
CHARGING PROCESS, AND THE VOLTAGE IS PRESENT IN
THE MAINS, IT MAY BE DUE TO A DEFECTIVE BATTERY.
IN THIS CASE, THE UNIT TURNS OFF AUTOMATICALLY TO
AVOID
OVERHEATING
CONTACT TO THE TECHNICAL SERVICE.
The Unit can operate in Stand-Alone mode, that is, operating
without mains being present or unplugged from mains.
This button is used to turn ON the Collimator Lamp from the Control Panel.
The Lamp remains illuminated for a few seconds before automatically switching
off.
THE
REMAINING
0188K EN 20201116
BATTERIES.