FINE POSITIONING CONTROLS:
The four buttons on the Hand-grips control the motion of each driving wheel
(forwards / backwards). This permits fine positioning adjustment of the Unit
respecting the patient, with the operator positioned opposite the
Tube-Collimator Assembly.
Fine Positioning velocity is reduced as this control is not designed for
displacements.
The buttons correspond to each motor and do not change when the Unit is in
Parking Position.
After eight seconds pressing any of the buttons (Fine
Positioning Controls) on the Hand-grips, the Unit stops
moving. Release these buttons and press them again to
enable the Unit displacement.
FOR THE CORRECT OPERATION OF THE X-RAY MOBILE
UNIT, THE USER MUST HAVE DRY HANDS WHEN WORKING
WITH THE SYSTEM.
DO NOT USE OR DRIVE THE SYSTEM WITH WET HANDS OR
IMPREGNATED WITH DISINFECTANT GEL OR ANY OTHER
SUBSTANCE OR LIQUID, SPECIALLY WHEN USING THE
MOVEMENT CONTROLS (HANDLEBAR, HEAD-ASSEMBLY
HANDGRIPS); OTHERWISE, THESE SUBSTANCES COULD
CAUSE SYSTEM MALFUNCTION AND/OR AN INCORRECT
OPERATION OF THE MOTION CONTROLS.
IN THIS CASE, TURN OFF THE UNIT AND CLEAN THE
AFFECTED PARTS.
In order to avoid uncontrolled displacement of the Unit
during the Start-up, due to a failure of the displacement
controls
(Fine
short-circuited),
commands are blocked although the unit can be controlled
with the Handlebar.
The unit displacement can also be blocked during the driving.
An audio signal is emitted (beep sequence in 2 seconds
intervals) to alert the user about a failure condition (refer to
Table 3-1).
DX-D 100 Mobile X-ray Unit
Positioning
Controls
movements
controlled
User Manual
pressed
or
with
these
55