GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or the user of this X-ray System should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000--4--5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 61000--4--11
Power frequency (50/60 Hz)
magnetic field
IEC 61000--4--8
NOTE - - U
is the a.c. mains voltage prior to application of the test level.
T
0188K EN 20201116
(IEC 60601-1-2:2007)
IEC 60601-1-2:2007
Test Level
6 kV contact
¦
8 kV air
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
1 kV line(s) to line(s)
¦
2 kV line(s) to earth
¦
< 5 % U
T
(> 95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
< 5 % U
T
(> 95 % dip in U
)
T
for 5s
3 A/m
Compliance
Electromagnetic environment - - guidance
level
6 kV
Floors should be wood, concrete or ceramic tile.
¦
If floors are covered with synthetic material, the
8 kV
relative humidity should be at least 30%.
¦
2 kV
¦
Mains power quality should be that of a typical
commercial or hospital environment.
1 kV
¦
1 kV
¦
Mains power quality should be that of a typical
commercial or hospital environment.
2 kV
¦
> 95 %
for 0.5 periods
Mains power quality should be that of a typical
60 %
commercial or hospital environment. If the user
for 5 periods
of the X-ray System requires continued
operation during power mains interruptions, it is
recommended that the X-ray System be
powered from a uninterruptible power supply or
30 %
a battery.
for 25 periods
> 95 %
for 250 periods
Power frequency magnetic fields should be at
3 A/m
levels characteristic of a typical location in a
(50 Hz)
typical commercial or hospital environment.
DX-D 100 Mobile X-ray Unit
User Manual
29