AGFA DX-D 100 User Manual page 70

Mobile x-ray unit
Hide thumbs Also See for DX-D 100:
Table of Contents

Advertisement

DX-D 100 Mobile X-ray Unit
User Manual
3.10 DOSIMETRY (OPTIONAL)
.
Note
Tab
62
The optional Radiation Meter is installed under the Collimator and reads
radiation as Dose Area Product (DAP) in mGy*cm
Radiation Meter ).
Do not install any accessories between the Radiation Meter and
the patient. This will disturb the radiation reading.
The Radiation Meter can be removed from the rail system to be cleaned or
serviced. To remove the Radiation Meter, pull back on the two tabs which lock
the Radiation Meter to the rails and pull out the Radiation Meter. The cable
connection to the PC is found behind the Collimator.
2
(refer to the manual of the
Radiation Meter
(optional)
0188K EN 20201116