Removing The High Voltage Module Assembly - ZOLL M Series Service Manual

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual
7. Removing the
High Voltage
Module Assembly
Tools Required
• #2 Phillips screwdriver
• Exacto-Knife
WARNING! This unit may contain lethal
voltages. You must completely discharge the
high voltage capacitor by changing the energy
selection during the charge cycle of the
defibrillator. Power off unit. Wait three minutes
before disassembly.
DO NOT SHORT THE TERMINAL ENDS OF
THE CAPACITOR. Refer to Section 8, Step 2.
Setup
1. Remove the Front Panel Assembly.
2. Remove the Upper Housing Assembly.
3. Remove the System Board.
4. Remove the Battery Interconnect Board.
5. Remove the High Voltage/Charger
Assembly
To reinstall the High Voltage Module Assembly,
reverse the steps.
Step 1: Remove the signal cable from the High
Voltage module.
.
Step 3: Separate the High Voltage Module from the
foam.
77
Step 2: Remove the Kapton tape on the bottom of
the foam surrounding the High Voltage Module.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents