M Series Service Manual
Do this...
2.18
If applicable, connect D/C
current and install the
battery. Turn the unit off.
2.19
Remove the battery.
2.20
Replace the battery and the
turn unit on.
2.21
Press the ANALYZE
button (if available).
2.22
Move the selector switch to
DEFIB. Select 2J. Press the
CHARGE button.
2.23
Press and hold the
ENERGY SELECT down
arrow.
2.24
Press and release the
ENERGY SELECT up
arrow 19 times.
2.25
Press the CHARGE button.
2.26
Press the SHOCK button.
Record your results on the Maintenance Test Checklist.
Observe this...
CHARGER ON indicator lights.
The amber or green lights illuminate.
The yellow light indicates the battery is being charged. The green
light indicates the battery is fully charged to present capacity.
Note: If both lights flash ON/OFF, the unit is defective or no
battery is installed.
Note that both charge lights (green and amber) flash alternately.
Note that the yellow charge light illuminates.
SELECT DEFIB MODE message appears on the monitor. (For
manual devices.)
The display shows that the unit is charging. The SHOCK button
lights when the unit is charged. Ready tone for DEFIB sounds.
Unit discharges internally and selected energy decrements to 1J.
The following energy amounts display incrementally 2, 3, 4, 5, 6,
7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 360J (for DSW).
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, 200J.
Note the display shows the unit charged up to 360J (200J -
Biphasic) and the SHOCK button lights.
The unit discharges and the SHOCK button is no longer lit. A 15
second strip chart automatically prints, displaying the number of
joules delivered (if equipped with recorder).
9
Pass/Fail/NA