WARNING!
CAUTION TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU CONDUCT
DEFIBRILLATOR TESTS OR REPAIRS.

Safety Precautions You Need to Take

SHOCK HAZARD!
• Only properly trained technicians should service the unit.
• The unit can contain deadly voltages even if the unit is turned off.
• Make sure to discharge the unit before working with it.
• Make sure you take the necessary precautions when working with static sensitive units. For example, you must wear
a conductive wrist strap (which touches your skin) connected to a grounding mat and to the earth ground. You must
remove the wrist strap when you discharge high voltage or when you are working on energized equipment.