11.0 Defibrillator Self Test
SHOCK HAZARD!
TAKE THE NECESSRY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU
START CONDUCTING THE DEFIRBILLATOR TESTS.
Keep hands and all other objects clear of the multi-function cable connections and defibrillator analyzer when
discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.
CAUTION
Do NOT internally discharge the unit more than 3 times in 1 minute. Note that multiple rapidly repeating internal
discharges at more than 30 Joules may damage the unit.
MFC Test Port Connector 1004-0053-99 with universal cable.
Tools Needed
MFC Test Adaptor Connector (Dynatech Nevada Part Number 3010-0378 or equivalent).
Dynatech Impulse 4000 or equivalent defibrillator analyzer.
ECG Cable.
Stop watch.
Ensure the unit is turned off and the ECG cable is connected to the unit and analyzer. The universal cable should not
Test Setup
be connected to any equipment at the beginning of this test.
Do this...
11.1
Turn the selector switch to
DEFIB (ON for AED.
Select Manual mode.) Set
leads to PADS.
Observe this...
CHECK PADS/POOR PAD CONTACT message displays.
23
Pass/Fail