M Series Service Manual
Error Message
Explanation
DEFIB FAULT 86
One discharge switch closed during power up test.
DEFIB FAULT 87
Both discharge switches closed during power up test.
DEFIB FAULT 94
Processor fault causing safety monitor port to be
non-functional.
DEFIB FAULT 95
Safe or shutdown line is not functional.
DEFIB FAULT 96
XPATREL or XPAT_ENABLE is faulted or one of
the discharge transistors has shorted.
DEFIB FAULT 108
VMON voltage is less than the target energy during
charging.
DEFIB FAULT 109
Defib capacitor voltage is greater than selected
energy when defibrillator is charging or ready.
DEFIB FAULT 111
Defib capacitor voltage has exceeded the absolute
maximum acceptable voltage.
DEFIB NOT CHARGED
Discharge button is pressed but the unit is not
charged.
50
User
Technical Action
Advisory
Replace paddles,
control board or system
board.
Replace paddles,
control board or system
board.
Replace system board,
high voltage module or
capacitor.
Replace high voltage
module.
Replace high voltage
module.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module, capacitor, and
or system board.