M Series Service Manual
Error Message
Explanation
CHECK RECORDER
Produced when paper tray is empty, paper jams or
recorder door is opened.
CHECK SPO
SITE
Low or no perfusion in monitored finger or toe.
2
CHECK SPO
SENSOR
Reposition SpO
2
CLOCK FAULT 11
Real time clock oscillator failure.
CLOCK FAULT 12
Real time clock back-up power supply failure. Found
oscillator stopped at power-up, but oscillator now
running when the system is running. (Oscillator only
runs when main power is applied).
CLOCK FAULT 13
One of the set time units (seconds, minutes, year,
etc.) is out of range.
CO
COMM ERROR
No or invalid communication from the EtCO
2
module.
CO
SENSOR WARM UP
EtCO
2
CONFIRM MANUAL
Displayed when manual mode is entered. Alerts user
MODE
to confirm that manual mode is desired.
DEFIB DISABLED
User prompt issued simultaneously with other faults
if defib is disabled.
sensor on patient.
2
Sensor warming up.
2
48
User
Technical Action
Advisory
Replace paper sensor
board, system
interconnect board,
and/or system board.
Replace system board.
Replace system board.
Replace system board.
Replace EtCO
2
and or system board.
Wait for sensor to warm
up. This process takes
up to approximately
one minute.
Possible configuration
problem. Replace high
voltage module. Call
ZOLL Technical
Support.
module
2