M Series Service Manual
Error Message
Explanation
DEFIB FAULT 71
More than 50 internal dumps occurred in less than 20
minutes.
DEFIB FAULT 72
General defib error.
DEFIB FAULT 76
Capacitor voltage too high for selected energy.
DEFIB FAULT 77
Capacitor voltage > than absolute rated max.
DEFIB FAULT 78
Unable to charge defib cap.
DEFIB FAULT 79
Defibrillator charging too slowly.
DEFIB FAULT 80
4 defibrillator faults detected within 20 second
period.
DEFIB FAULT 81
Discharge switch in undefined state.
DEFIB FAULT 84
"Upper" discharge transistor shorted (measured via
applicable A/D channel).
DEFIB FAULT 85
"Lower" discharge transistor shorted (measured via
applicable A/D channel).
49
User
Technical Action
Advisory
Turn the unit to OFF
and back on. If fault
persists, replace high
voltage module.
Turn the unit to OFF
and back on. If fault
persists, replace high
voltage module.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module.
Replace high voltage
module.