M Series Service Manual
Do this...
18.3
On the Novametrix
Simulator, set SENSOR
LOCATION to REF
CELL.
18.4
On the Novametrix
Simulator, set SENSOR
LOCATION to AA CELL.
Set% CO
to 10.
2
Set CO
mode to
2
RESPIRATION.
18.5
On the Novametrix
Simulator, set% CO
to 5.
2
Record your results on the Maintenance Tests Checklist.
Observe this...
The EtCO
reading of 36-40 mmHg displays on the monitor.
2
Note that you may need to wait up to 10 seconds for the unit to
stabilize.
The EtCO
reading of 74-84 mmHg appears on the M Series
2
display. Note that you may need to wait up to 10 seconds for the
unit to stabilize.
The EtCO
reading of 34-42 mmHg displays on the M Series
2
monitor.
Note that you may need to wait up to 10 seconds for the unit to
stabilize. The Respiration Rate (RR) of 22-24 displays on the M
Series monitor. Press WAVE 2 softkey. The EtCO
displays. Press RECORDER button. The EtCO
Note that the CO
waveform is displayed and printed at 12.5
2
millimeters per second scale.
37
Pass/Fail
waveform
2
waveform prints.
2