6. Removing the
High Voltage/
Charger Assembly
(Continued)
Step 5: Lift the SpO
module straight up from the
2
foam and disconnect the ribbon cable from the
Isolated Power Supply.
Step 7: (Biphasic units only) Disconnect the Molex
connector from the biphasic module.
Molex connector
biphasic module
76
Step 6: (Biphasic units only) Disconnect the biphasic
defibrillator capacitor from the rear panel connector.
biphasic defibrillator capacitor
Step 8: (EtCO
units only) Remove the EtCO
2
connector by gently lifting the connector away from
the back panel. To remove the EtCO
power supply, refer to Step 5.
EtCO
connector
2
2