M Series Service Manual
Error Message
Explanation
RECORDER FAULT 143
Strip chart failed power-up echo test.
Communications error.
RECORDER FAULT 147
Strip chart printhead over temperature.
RELEASE BUTTONS
Simultaneous external paddle button presses
detected before unit reached full defib charge (ready
state).
RELEASE SHOCK
Discharge switch(es) closed when pressing charge
button. Discharge button pressed before defib
reached ready state.
REPLACE BATTERY
Battery voltage < absolute min. Shutdown imminent.
REPLACE EtCO
EtC0
2
SENSOR
more than five minutes. Sensor defective.
REPORT FULL
Summary report memory full.
REPORT HALTED
Summary report stops printing unexpectedly.
SENSOR WARM UP message displays for
2
56
User
Technical Action
Advisory
Check paper tray and
paper path. Replace the
system interconnect
board and/or the system
board. Turn unit off and
back on again.
Check paper tray and
paper path. Replace the
print head, system
interconnect board and/
or the system board.
Release buttons.
Release shock button.
Check paddles.
Replace controls board.
Replace with charged
battery.
Replace sensor cable.
Erase summary report.
Turn unit off and then
back on again. Print
Summary again. If fault
persists, replace system
board.