Temperature-Resistance Thermometer Calibration - Omron K3NH Operation Manual

Temperature meter
Table of Contents

Advertisement

Maintenance Mode

5-3-5 Temperature-resistance Thermometer Calibration

Preparation
Calibration
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr
13. Press the Shift Key to select whether or not to accept the settings.
14. Press the Up Key to select yes.
15. Press the Mode Key to overwrite all calibration data and to put the K3NH into
RUN mode. The calibration data will not be overwritten if no is selected.
• Make sure that the thickness of each connected wire is the same.
• Dial 6 in the following diagram represents a high-precision resistance box.
• Short-circuit terminals 3 and 4 with each other.
1. First, the sure alarm display will appear. This will not appear once field cal-
1, 2, 3...
ibration is performed.
2. Press the Mode Key to display p300. The teaching indicator will be lit. Then
calibrate the main input.
Set Value LED Display Model
Set Value LED Display Model
RUN mode
Dial 6
Set Value LED Display Model
Set Value LED Display Model
Section 5-3
Basic Model
Basic Model
100 to 240 VAC
12 to 24 VDC
Basic Model
Basic Model
85