Maintenance Mode
Procedure for Field Calibration
5-3-4 Thermocouple Calibration
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr
• Only input set with the input type parameter in setup mode can be the object of
input calibration. Press the Mode Key for approximately one second to tenta-
tively register the parameter data of all calibration items.
• The data registration setting str does not appear unless all calibration items
are tentatively registered.
1, 2, 3...
1. Press the Mode and Shift Keys simultaneously while turning the K3NH on to
display mant maintenance mode setting.
2. Press the Mode Key for more than one second to display re in set value
initialization setting.
3. Press the Mode Key to display the calib field calibration setting.
4. Press the Shift Key to display no.
5. Press the Up Key to display yes. Press the Mode Key to select the calibra-
tion item.
• There are two calibration methods. One applies to thermocouple group 1 (i.e.,
K1, J1, L1, E, N, W, and PLII) and the other applies to thermocouple group 2
(K2, J2, L2, R, S, B, T, and U).
Set Value LED Display Model
Set Value LED Display Model
Set Value LED Display Model
Set Value LED Display Model
Set Value LED Display Model
User calibration
Section 5-3
Basic Model
Basic Model
Basic Model
Basic Model
Basic Model
79