Maintenance Mode
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr
8. The display will change to pb 0 for 0-Ω calibration. Short-circuit terminals 2,
3, and 4 to one another. The teaching indicator will be lit.
Set Value LED Display Model
Short-circuit terminals 2 to 4.
9. Press the RESET/TEACH Key to display the calibration value setting. The
teaching indicator will flash. After the value stabilizes (i.e., the fluctuation of
the value is less than several digits), press the Mode Key to tentatively regis-
ter the calibration data.
Set Value LED Display Model
10. The display will change to str for confirmation of the settings.
Set Value LED Display Model
11. Press the Shift Key to select whether or not to accept the settings.
Set Value LED Display Model
12. Press the Up Key to select yes.
13. Press the Mode Key to overwrite all calibration data and to put the K3NH into
RUN mode. The calibration data will not be overwritten if no is selected.
Set Value LED Display Model
Basic Model
Basic Model
Basic Model
Basic Model
Basic Model
RUN mode
Section 5-3
87