Maintenance Mode
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr
84
the value stabilizes (i.e., the fluctuation of the value is less than several dig-
its), press the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
8. The display will change to k 0 for 0-mV calibration. The teaching indicator
will be lit.
Set Value LED Display Model
9. Press the RESET/TEACH Key to display the calibration value setting. Then
set the output of the STV to 0 mV. The teaching indicator will flash. After the
value stabilizes (i.e., the fluctuation of the value is less than several digits),
press the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
10. Calibrate the bias compensation value as the last item. The display will
change to bias. The teaching indicator will be lit. Disconnect the STV, set
the cold junction compensator to 0°C, and enable the built-in thermocouple.
Be sure to disconnect the wires connected to the STV.
Set Value LED Display Model
11. Press the RESET/TEACH Key to display the calibration value setting. The
teaching indicator will flash. After the value stabilizes (i.e., the fluctuation of
the value is less than several digits), press the Mode Key to tentatively regis-
ter the calibration data.
Set Value LED Display Model
12. The display will change to str for confirmation of the settings.
Set Value LED Display Model
Section 5-3
Basic Model
Basic Model
Basic Model
Basic Model
Basic Model
Basic Model