Testing Performance
To teslthe threshold accuracy (logic analyzer)
Test the next pod
1 Using the 17 -by-2 test connector and probe tip assembly, connect the data and clock
channels of the next pod to the output of the function generator until all pods have
been tested.
To unassign a pod
pair
and assign the next pod pair to be tested, press the Config key. Select
the pod pairs, then select assign or unassign in the pop-up menu.
2 Start with "Test the TTL threshold" on page 3-11, substituting the next pod to be
tested for pod 1.
3-17